Overload protection
A input: 10 A (500 V) fast blow fuse
µA input: 600 V rms
*1
AC conversion type:
Conversion type and additional specification are the same as DC/AC voltage
Resistance
Range Accuracy Overload protection
600.0 Ω *2
6.000 KΩ
60.00 KΩ
600.0 KΩ
6.000 MΩ ± (1.0% + 2 digits)
60.00 MΩ *1 ± (1.5% + 2 digits)
Open circuit voltage: -1.3 V approx.
*1 <100 digit rolling
*2 <10 digit rolling
Diode check and continuity
Range Resolution Accuracy
10 mV ± (1.5% + digits*)
* For 0.4 V ~ 0.8 V
Max. test current: 1.5 mA
Max. open circuit voltage: 3 V
Overload protection: 600 V rms
Continuity: Built-in buzzer will sound when the
resistance is less than 500 Ω approx.
Response time is 100 ms approx.
± (0.7% + 2 digits)
600 V rms